Back to finder
7866+7872D Hydrogen Purity Analyzer
H2
OnlineSampling systemExplosion-proof

Product no. 7866+7872D

7866+7872D Hydrogen Purity Analyzer

Online hydrogen purity analysis for generator hydrogen cooling systems and gas replacement monitoring in power plant applications.

Gas type
Hydrogen
Product group
H2 Gas analyzers
Application
Generator hydrogen purity analysis

Key Product Data

Measurement range
0 to 100 % CO2 in air; 0 to 100 % H2 in CO2; 75 to 100 % H2 in air
Accuracy
Plus or minus 2% F.S under standard conditions
H2 response time
63%: 13 s; 90%: 23 s
CO2 response time
63%: 24 s; 90%: 45 s
Output
4 to 20 mA, factory setting 75 to 100 % H2 in air
Explosion-proof rating
Ex d IIC T6
Sample gas flow
0.5 SCFH
Model
7866+7872D
Online hydrogen purity analyzer with sampling system
Covers CO2 in air, H2 in CO2, and H2 in air measurement stages
Suitable for generator hydrogen cooling and gas replacement monitoring
Back to catalogue

Product Overview

Online hydrogen purity analysis for generator hydrogen cooling systems and gas replacement monitoring in power plant applications.

Product Advantages

  • Online hydrogen purity analyzer with sampling system
  • Covers CO2 in air, H2 in CO2, and H2 in air measurement stages
  • Suitable for generator hydrogen cooling and gas replacement monitoring

Technical Specifications

12 parameters
Measurement range
0 to 100 % CO2 in air; 0 to 100 % H2 in CO2; 75 to 100 % H2 in air
Accuracy
Plus or minus 2% F.S under standard conditions
H2 response time
63%: 13 s; 90%: 23 s
CO2 response time
63%: 24 s; 90%: 45 s
Output
4 to 20 mA, factory setting 75 to 100 % H2 in air
Explosion-proof rating
Ex d IIC T6
Sample gas flow
0.5 SCFH
Model
7866+7872D
Type
Online hydrogen purity analyzer with sampling system
Dimensions
800 mm x 400 mm x 1660 mm
Weight
131 kg
Maximum load
800 ohm

Application and Selection Checklist

This product is listed for generator hydrogen purity analysis. Before selection, provide the following project information so the configuration can be reviewed against the actual operating conditions.

  • Gas composition and operating conditions
  • Required range, capacity, flow, or response
  • Power supply, interfaces, and installation space
  • Site environment and destination country
  • Quantity, schedule, and operating workflow
  • Required standards, drawings, tests, and certificates

Standards, Documentation, and Project Fit

Product suitability is determined by the selected model, configuration, operating conditions, interfaces, and destination requirements. A matching product in the same group is not automatically interchangeable.

Standards, declarations, calibration records, inspection documents, drawings, manuals, commissioning, training, and test scope vary by product and contract. State every required item in the inquiry and confirm it in writing before purchase.

Typical application descriptions on this page are selection references, not claims about a named customer installation. Final project fit requires review of the actual site data.

Frequently Asked Questions

What is 7866+7872D Hydrogen Purity Analyzer used for?

7866+7872D Hydrogen Purity Analyzer is intended for generator hydrogen purity analysis. Online hydrogen purity analysis for generator hydrogen cooling systems and gas replacement monitoring in power plant applications.

Which product group does 7866+7872D belong to?

7866+7872D is listed in the H2 Gas analyzers product group within the Hydrogen range.

What should be confirmed before selecting 7866+7872D?

Confirm the application, gas conditions, required performance, utilities, connection interfaces, site environment, quantity, destination, and required technical documents.

Are standards and certificates available for 7866+7872D?

Applicable standards, declarations, calibration or inspection records, and certificate availability depend on the selected model and destination. Confirm required documents in writing during the inquiry.

Related Selection Guides and Products

Product Images

1 image
7866+7872D Hydrogen Purity Analyzer product image 1